Page: 252 |
Affects: W.F. spike cross-correlation histogram
Each action potential on the waveform selected by the "Spike W.F. #", in the range to be analysed, triggers a "sweep" from the waveform selected by the "Spike corr. W.F. #". The window and delay for these sweeps are set as they are for spike triggered averages. Each sweep window is itself scanned for spikes, so you have to set the waveform parameters for spike detection for both of the above waveforms. These parameters are a part of the waveform parameter files associated with each waveform.
The resulting graph shows the average number of action potentials on this waveform (selected by the "Spike corr. W.F. #"), corresponding to time intervals before or after spikes on the other waveform.
The parameters "Corr. spikes before trigger" and "Corr. spikes after trigger" limit the number of spikes counted in each sweep.
Key sequence | Parameter | Initial Value |
---|---|---|
<Esc>SSCN | Spike corr. W.F. # | ? |